Hyperthermia has long been known as a radiation therapy sensitizer of high potential; however successful delivery of this modality and integrating it with radiation have often proved technically difficult. We present the dual-modality thermo-brachytherapy (TB) seed, based on the ubiquitous low dose-rate (LDR) brachytherapy permanent implant, as a simple and effective combination of hyperthermia and radiation therapy. Heat is generated from a ferromagnetic or ferrimagnetic core within the seed, which produces Joule heating by eddy currents. A strategically-selected Curie temperature provides thermal self-regulation. In order to obtain a uniform and sufficiently high temperature distribution, additional hyperthermia-only (HT-only) seeds are proposed to be used in vacant spots within the needles used to implant the TB seeds; this permits a high seed density without the use of additional needles.
Experimental and computational studies were done both to optimize the design of the TB and HT-only seeds and to quantitatively assess their ability to heat and irradiate defined, patient-specific targets. Experiments were performed with seed-sized ferromagnetic samples in tissue-mimicking phantoms heated by an industrial induction heater. The magnetic and thermal properties of the seeds were studied computationally in the finite element analysis (FEA) solver COMSOL Multiphysics, modelling realistic patient-specific seed distributions. This distributions were derived from LDR permanent prostate implants previously conducted at our institution; various modifications of the seeds' design were studied. The calculated temperature distributions were analyzed by generating temperature-volume histograms, which were used to quantify coverage and temperature homogeneity for a range of blood perfusion rates, as well as for a range of seed Curie temperatures and thermal power production rates. The impact of the interseed attenuation and scatter (ISA) effect on radiation dose distributions of this seed was also quantified by Monte Carlo studies in the software package MCNP5.
Experimental and computational analyses agree that the proposed seeds may heat a defined target with safe and attainable seed spacing and magnetic field parameters. These studies also point to the use of a ferrite-based ferrimagnetic core within the seeds, a design that would deliver hyperthermia of acceptable quality even for the high rate of blood perfusion in prostate tissue. The loss of radiation coverage due to the ISA effect of distributions of TB and HT-only seeds may be rectified by slightly increasing the prescribed dose in standard dose superposition-based treatment planning software.
A systematic approach of combining LDR prostate brachytherapy with hyperthermia is thus described, and its ability to provide sufficient and uniform temperature distributions in realistic patient-specific implants evaluated. Potential improvements to the previously reported TB seed design are discussed based on quantitative evaluation of its operation and performance.